| Text of Test Method 7E - NOx - Instrumental - Posting Date: 11/6/2008 |
 |
FREQUENTLY ASKED QUESTIONS (FAQS)
- Does the new interference test apply to analyzers
that have already met the requirements of the old
Method 6C, 7E, 10, and 20 interference tests?
- Section 8.2.7(2) implies that the interference test
needs to be repeated after replacement of a major piece
of equipment. Does this mean a like kind replacement?
- Specifically, which analyzers are required to conduct
the Manufacturer Stability Test (MST)?
- Is the Manufacturer Stability Test (MST) performed by
the manufacturer, the tester, or either?
- Would the MST requirements apply for multiple range instruments
that are occasionally used with a calibration span of <20 ppm?
- The revisions to Method 7E require an NO2 converter efficiency
check before each field test. If we are at a facility for 4 days,
and we test 4 separate sources for NOx using the same instrument
over the 4-day period, is that considered a single field test of
that analyzer or is it considered 4 field tests? The language is
different for calibration error, where it says to conduct the
calibration error test prior to performing the first run.
Wouldn't the same language be used if a daily converter check
was intended?
In another situation, a large number of Part 60 and Part 75
RATAs are conducted each year; during some of the RATA test
programs, NOx emissions will be measured from 6-8 boilers in
as many consecutive days. In this instance, what is the
appropriate frequency for the conversion efficiency check:
prior to performing each RATA test set; at the beginning of
each test day, or only at the beginning of the multi-day test program?
- Method 7E requires that the converter efficiency test for dilution
systems be completed at a point in the sample system prior to
dilution. Can this be changed to allow for direct injection to
the analyzer, as per non-dilution systems?
- Section 10.0 requires that analyzers be calibrated for all species
of N0x that they detect. If your analyzer measures N0 and N02
separately, then you must use N0 and N02 calibration gases.
Can an N0 protocol gas (calibration gas that is certified for N02)
be used for calibration if the N02 concentration is on the certification?
There is a limited supply of NO2 calibration gas; however, one can obtain
an analysis for total N0x which will give a good indication of this
certified value.
- Currently, the lowest EPA protocol calibration gases we are aware
that can be made are:
1. S02 - 5 ppm
2. N0 - 1 ppm
3. C0 - 1 ppm
4. C02 - 1000 ppm
5. 02 - 0.2%
The problem is that NIST currently does not have a zero standard
from which we are allowed to create a zero protocol as required
under the requirements to create a protocol gas. It is important
for the testers to understand that the ranges listed above are the
lowest ranges in EPA Protocol gases that are available. A tester
could order a non-protocol zero gas to use as the low range (0-20%)
calibration gas and find that the test is not valid because the gas
is not of the appropriate quality.
- Please confirm my understanding of the new calibration concentrations:
High Level Gas = 20-100 percent of the calibration span
Mid Level Gas = 40-60 percent of the calibration span
Low Level Gas = Less than 20 percent of the calibration span
- Method 7E specifies an NO2 gas of 40-60 ppmv for converter efficiency
tests. Would it be allowable to use a lower concentration on a
case-by-case basis?
- Method 7E states that sample flow rate must be maintained
within 10 percent of the flow rate used during system bias
tests, and that sample components must be maintained below
the dew point of the gas. Will testers be required to document
these variables?
- Method 7E requires that the particulate filter be included
in system bias tests. Will using an in-stack filter and
flood-the-probe approach to performing system bias tests be
allowed?
- Is the stratification test specified in Method 7E required
for RATAs?
- Should Equation 7E-4 be divided by CS and multiplied by 100?
- Equation 7E-5 only works when a zero gas is used as the
low-level calibration gas. Can the equation in old Method 3C
be used when a non-zero calibration gas is used?
- When we use new Equation 7E-9, we obtain results that are
inordinately high. Is there an error in this equation?
- We do not understand the purpose of Paragraphs 16.2.2(2) and (3).
Can you clarify their intent?
- Do the test gases in the interference test have to be NIST traceable?
- Are ambient-level analyzers used in conjunction with dilution probes
required to pass the manufacturer's stability test?
- If you have a source that does not have emissions above 4 ppm for CO
and 4 ppm for SO2, how do you pick the calibration span if you cannot
obtain gases at that low concentration? The old standard would give you
the option of picking the span based on the emission standard but I do not
see that in the new requirements.
- Section 8.1.1 of Method 7E requires a stratification test consisting of 12
or 3 points to determine the number of sampling points needed for sample
collection. Is this practical for small-stack facilities such as stationary
internal combustion engines, especially since most are equipped with a
muffler that enhances gas mixing?
- If testing for multiple pollutants or diluents at a site, is a stratification
test required for each pollutant/diluent?
Does the new interference test apply to analyzers that have
already met the requirements of the old Method 6C, 7E, 10,
and 20 interference tests?
No. Current instruments that will have already passed
the interference test requirements under the old methods
before August 14, 2006 are grandfathered from the new
interference test requirements.
Section 8.2.7(2) implies that the interference test needs
to be repeated after replacement of a major piece of equipment.
Does this mean a like kind replacement?
This does not mean original replacement or like kind parts.
The interference test must be repeated only when a different
model replacement part is used and the effects of potential
interferences on this part are unknown.
Specifically, which analyzers are required to conduct the
Manufacturer Stability Test (MST)?
Only those analyzers that are dedicated to low-concentration
(<20 ppm) measurements are required to conduct the MST.
Is the Manufacturer Stability Test (MST) performed by the
manufacturer, the tester, or either?
The MST would primarily be performed by the manufacturer;
however, it could be performed by the tester. The final rule
reference to 40 CFR 53.55 and 53.56 were in error. Only the
provisions in 40 CFR 53.23 for temperature and line voltage
stability apply for the MST. The noted temperatures and line
voltages in 53.23 should be modified, as necessary, to mimic
the anticipated field conditions for the analyzer.
Would the MST requirements apply for multiple range instruments
that are occasionally used with a calibration span of <20 ppm?
No.
The revisions to Method 7E require an NO2 converter efficiency
check before each field test. If we are at a facility for 4 days,
and we test 4 separate sources for NOx using the same instrument
over the 4-day period, is that considered a single field test of
that analyzer or is it considered 4 field tests? The language is
different for calibration error, where it says to conduct the
calibration error test prior to performing the first run.
Wouldn't the same language be used if a daily converter check
was intended?
In another situation, a large number of Part 60 and Part 75
RATAs are conducted each year; during some of the RATA test
programs, NOx emissions will be measured from 6-8 boilers in
as many consecutive days. In this instance, what is the
appropriate frequency for the conversion efficiency check:
prior to performing each RATA test set; at the beginning of
each test day, or only at the beginning of the multi-day test program?
A fixed definition for "field test" might prove too restrictive.
We recommend conducting the converter efficiency test before each
test day. However, if one is willing to take the risk, an efficiency
test at the end of several days of testing is acceptable if the final
efficiency check passes. If such a final test is failed, the data is
invalidated beyond the last day the last converter test was passed.
Method 7E requires that the converter efficiency test for dilution
systems be completed at a point in the sample system prior to
dilution. Can this be changed to allow for direct injection to
the analyzer, as per non-dilution systems?
Yes, this inadvertent requirement will be corrected in the future
correction notice.
Section 10.0 requires that analyzers be calibrated for all species
of N0x that they detect. If your analyzer measures N0 and N02 separately,
then you must use N0 and N02 calibration gases. Can an N0 protocol gas
(calibration gas that is certified for N02) be used for calibration if
the N02 concentration is on the certification? There is a limited supply
of NO2 calibration gas; however, one can obtain an analysis for total
N0x which will give a good indication of this certified value.
The Section 10 requirement to calibrate instruments that measure NO
and NO2 with NO and NO2 calibration gas was added primarily for those
analyzers that don't use an NO2 to NO converter. This requirement does
not apply to systems with converters. Any gas combination that has
certification and doesn't contain components that interfere with the gas
being detected is allowed. A gas certified for total NOx is acceptable
for calibrating instruments that measure NO and NO2.
Currently, the lowest EPA protocol calibration gases we are aware that
can be made are:
1. S02 - 5 ppm
2. N0 - 1 ppm
3. C0 - 1 ppm
4. C02 - 1000 ppm
5. 02 - 0.2%
The problem is that NIST currently does not have a zero standard from
which we are allowed to create a zero protocol as required under the
requirements to create a protocol gas. It is important for the testers
to understand that the ranges listed above are the lowest ranges in EPA
Protocol gases that are available. A tester could order a non-protocol
zero gas to use as the low range (0-20%) calibration gas and find that
the test is not valid because the gas is not of the appropriate quality.
In expanding the low-concentration calibration gas range from "zero gas"
to "low-level gas," we retained the allowance to use zero gas. Not
explaining that a zero gas did not have to be of traceability protocol
quality was an inadvertent omission. We will correct this in Method 7E
and require that zero gas meet the requirements for "zero material" in
40 CFR 72.2. These zero gas requirements limit:
NOx, SO2, or THC < 0.1
CO < 1 ppm
CO2 < 400 ppm
Please confirm my understanding of the new calibration concentrations:
High Level Gas = 20-100 percent of the calibration span
Mid Level Gas = 40-60 percent of the calibration span
Low Level Gas = Less than 20 percent of the calibration span
Only your understanding of the high-level gas is incorrect.
The high-level gas concentration is equivalent to and sets
the calibration span. Section 7.1.1 directs the high-level
gas be chosen such that the emission measurements fall between
20 to 100 percent of this high calibration gas (which is the same
as the calibration span). The mid-level gas will then be 40-60%
of this high-level concentration (or calibration span) and the
low-level gas will be less than 20% of this high-level concentration
(or calibration span). One may choose a zero gas as the low-level gas.
Method 7E specifies an NO2 gas of 40-60 ppmv for converter efficiency
tests. Would it be allowable to use a lower concentration on a
case-by-case basis?
Yes. Approval would be required on a case-by-case basis.
Method 7E states that sample flow rate must be maintained
within 10 percent of the flow rate used during system bias
tests, and that sample components must be maintained below
the dew point of the gas. Will testers be required to document
these variables?
Documentation is recommended.
Method 7E requires that the particulate filter be included
in system bias tests. Will using an in-stack filter and
flood-the-probe approach to performing system bias tests be
allowed?
Yes, as long as the filter is made of a non-reactive material.
Is the stratification test specified in Method 7E required
for RATAs?
No. The stratification test in Method 7E does not take precedence
over the existing stratification test requirements in 40 CFR 60,
Appendix B and 40 CFR 75, Appendix A.
Should Equation 7E-4 be divided by CS and multiplied by 100?
No, it is already expressed as a percentage of CS.
Equation 7E-5 only works when a zero gas is used as the low-level
calibration gas. Can the equation in old Method 3C be used when
a non-zero calibration gas is used?
Yes. Equation 3A-1 in old Method 3A should be used when a
non-zero low-level calibration gas is used. This equation
will be added to Method 7E in the correction notice.
When we use new Equation 7E-9, we obtain results that are
inordinately high. Is there an error in this equation?
Equation 7E-9 is incorrect. The equation should be a determination
of how NOx final differs from NOx peak after the 30-minute
converter evaluation period. The correct equation should be
Eff NO2 = NOx peak - NOx final / NOx peak x 100. The current
equation is a mistaken attempt to calculate the conversion
efficiency of the converter without knowing the concentration
of the test gas.
We do not understand the purpose of Paragraphs 16.2.2(2) and (3).
Can you clarify their intent?
Paragraphs 16.2.2(2) and (3) were inadvertent inclusions in the
rulemaking and do not make sense in the context of the converter
efficiency test. These two paragraphs should be ignored;
they will be deleted in the correction notice.
Do the test gases in the interference test have to be NIST traceable?
No.
Are ambient-level analyzers used in conjunction with dilution probes
required to pass the manufacturer's stability test?
No. Ambient analyzers are grandfathered from this requirement.
They are already designed to meet the equivalent method requirements
of 40 CFR Part 53.
If you have a source that does not have emissions above 4 ppm for CO
and 4 ppm for SO2, how do you pick the calibration span if you cannot
obtain gases at that low concentration? The old standard would give you
the option of picking the span based on the emission standard but I do not
see that in the new requirements.
Section 3.4 of Method 7E defines the calibration span and states:
"To the extent practicable, the measured emissions should be between
20 to 100 percent of the selected calibration span." It doesn't appear
that this is practicable in your case. Therefore, you have the option
of choosing gases that are more applicable for your source. This means
gases that are reasonably available and at levels best suited to accomplish
the data quality objectives. If the objective is to show compliance with
an emission standard, then picking a calibration span gas based on the
emission standard is reasonable.
Section 8.1.1 of Method 7E requires a stratification test consisting of 12
or 3 points to determine the number of sampling points needed for sample
collection. Is this practical for small-stack facilities such as stationary
internal combustion engines, especially since most are equipped with a muffler
that enhances gas mixing?
The stratification test requirements in Method 7E do not lend themselves well
to the small-diameter stacks of stationary combustion engines. The emissions from
these sources should be well mixed after a muffler. Method 7E notes that stationary
internal combustion engines with stack diameters less than 4 inches are exempted from
these stratification requirements. We have also found that emissions from engines in
general are too temporally variable to render a stratification test meaningful. Therefore,
we are planning to revise Part 60 Subpart JJJJ and Part 63 Subpart ZZZZ to allow for
single-point tests. The revisions will be proposed in a rule entitled "Revisions to
Testing Methods and Testing Regulations" scheduled for publication in the Federal Register
by mid-year 2010.
If testing for multiple pollutants or diluents at a site, is a stratification test required for each pollutant/diluent?
A single stratification test using a single pollutant or diluent is all that is needed to verify
the stratification conditions at a source. You do not have to perform separate stratification
tests for each pollutant and diluent.
Still have questions on this method? Contact the EMC
expert Foston Curtis at
curtis.foston@epa.gov .
|