Alternate Test Procedures

Questions and Answers


You will need Adobe Reader to view some of the files on this page. See EPA's PDF page to learn more.

How do I write a method in EPA format?

The Guidelines and Format for Methods to be Proposed at 40 CFR part 136 or part 141 (PDF) (60 pp., 386 K) (the "Method Guidelines and Format"; July 1996) have been prepared to promote consistency among analytical methods and to streamline the method promulgation process.

Top of Page

How do I ask EPA to review my method under the CWA ATP Program?

Under the CWA part of the Alternate Testing Program (ATP), EPA has provided detailed guidance about the kind of information the Agency needs to evaluate methods for potential approval. Methods evaluated under the ATP program need to go through notice and comment rulemaking before they are incorporated into CWA regulations. You may find this information in three protocols: two for chemical methods, and one for microbiological methods:

  1. Protocol for EPA Approval of New Methods for Organic and Inorganic Analytes in Wastewater and Drinking Water (PDF) (61 pp., 236 K) March 1999
  2. Protocol for EPA Approval of Alternate Test Procedures for Organic and Inorganic Analytes in Wastewater and Drinking Water (PDF) (61 pp., 222 K) March 1999
  3. EPA Microbiological Alternate Test Procedure Protocol for Drinking Water, Ambient Water, and Wastewater Monitoring Methods (PDF) (95 pp., 586 K) April 2004

Note: The EPA does not have a protocol for toxicity testing under EPA's Whole Effluent Toxicity (WET) program.

Before submitting a method to EPA for an ATP review, a developer confers with EPA to design a method validation study. This study tests the new or modified method in several representative matrices and independent laboratories.

The method also must also be written in a standard format that includes all of the steps in an approved method, such as sample and data handling, and quality assurance requirements (Guidelines and Format for Methods to be Proposed at 40 CFR Part 136 or Part 141 (PDF) (60 pp., 386 K) July 1996).

Top of Page

Is a method that is reviewed under the CWA ATP program an approved method?

No, the method must be approved through rulemaking before the method may be added to 40 CFR Part 136 as an approved CWA method.

Top of Page

Are there any exceptions to rulemaking before use of an alternate test procedure?

Yes, but only for wastewater methods. Approval for nationwide use requires a rulemaking process. In the interim, a facility may apply to an EPA Region for a limited-use ATP approval letter, i.e. for use at that facility. Generally it is not necessary for the limited-use ATP applicant to submit data, or do a side-by-side comparison, if the method has already been reviewed under the CWA ATP program which requires multi-lab and comparability data.

Top of Page

Why does EPA have two validation protocols for chemical methods?

One protocol covers modifications to approved methods that require EPA review and subsequent approval through rulemaking. Validation procedures for these modified methods are described in the "ATP Method" validation protocol (PDF) (61 pp., 222 K).

The other protocol covers methods that use a new determinative technique, which usually means that the method uses a detector different from any detector used in existing approved methods. Validation procedures for these new methods are described in the "New Method" validation protocol (PDF) (61 pp., 333 K).

New methods and ATP methods (i.e., a modification to an approved method) must be approved through rulemaking before nationwide-use.

Top of Page

What is "136.6" Method Flexibility?

Many approved methods provide the flexibility to modify and use that method without prior approval of the modification. In 2007, EPA amended the wastewater regulations at 40 CFR 136.6 to describe additional (and to clarify existing) flexibility to modify any Part 136 chemical method without prior review.

See also Flexibility to Modify CWA Methods memo (PDF) (6 pp., 75 K; November 2007; About PDF).

Top of Page

Does 136.6 flexibility to modify methods only apply to CWA chemical methods?

Yes. This flexibility does not apply to drinking water methods, or to CWA microbiological or biological methods.

Top of Page

Where can I find the Guide to Method Flexibility document?

You can find the Guide to Method Flexibility here.